Thermo Scientific Sensititre AIM Instrucciones De Uso página 23

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 319
Service tools
The service icon is restricted for use by a Sensititre™ trained service engineer. There are no user
serviceable parts inside the AIM system.
Plate well offset setting
The performance of Sensititre AIM to inoculate standard 96-well microtitration plates (non-Sensititre
plates) has not been established. Although other brands of 96-well microtitration plates may be used on
the AIM instrument, they must be pre-qualified before use.
The plate well offset setting can be toggled through 3 settings by pressing the offset icon displayed
below. The instrument is configured for Sensititre plates that have a slightly different well offset compared
to standard 96 well microtiter plates. To accommodate both plate types there are several options.
Sensititre plates only with no offset.
(Default setting).
If the Sensititre and standard microtiter plates option is selected, an additional icon will be displayed on
the inoculation screen. This option allows switching between Sensititre and standard microtiter plates by
toggling the offset icon located on the left side of the screen.
Sensititre AIM™ Instructions for Use
The picture can't
Service tools icon
The picture can't
Standard microtiter plates only. Uses
required dosing offset.
Both Sensititre and standard
microtiter plates.
23