Other faults
Problem
Erratic dosing and
splashing
No response from the
Touchscreen
Sensititre AIM™ Instructions for Use
Pump Motion Error
Rotary Drive (rotary motion)
Pump Rotatory Alignment Error
Rotary Drive
Processor Error
Processor Error
Processor Error
Cause
Incorrect plate offset selected
Incorrect clamping of the dosing
head
Incorrectly loaded plate
Plate corner clamp not engaging
with plate correctly
Incorrect mechanical setup
No Power
Firmware problem
Physical damage
Ensure that the tube and dosing head are
correctly inserted into the instrument. Press
the touchscreen to reset the instrument.
If the issue persists, contact Service
Engineer.
Ensure that the pump mechanism rotary drive
is aligned vertically.
Press the touchscreen to reset the
instrument.
If the issue persists, contact Service
Engineer.
Press the touchscreen to reset the
instrument.
If the issue persists, contact Service
Engineer.
Press the touchscreen to reset the
instrument.
If the issue persists, contact Service
Engineer.
Press the touchscreen to reset the
instrument.
If the issue persists, contact Service
Engineer.
Solution
Check offset setting chosen is correct for the
required plate type.
Ensure the dosing head clamp is fully closed
during inoculation.
Check the plate is loaded with the correct
orientation.
Clean plate corner clamp with approved cleaning
solution. If unsuccessful, contact Service
Engineer.
Contact Service Engineer.
Ensure the instrument is powered on. (See Loss of
Power section).
Power cycle the instrument.
Contact Service Engineer.