A
The picture can't be
displayed.
C
A
Left facing blue arrow returns you to the previous screen.
Right facing blue arrow takes you to additional dosing patterns and onto the maintenance
B
and information screen.
C
Dosing patterns (See the Dosing pattern section for a detailed list of patterns).
7. Once the required dosing pattern has been selected, the screen will display the chosen pattern
along with the inoculum amount that will be delivered to the plate wells. The inoculum amount can
be changed by pressing the inoculum level displayed in the center of the screen.
The picture can't
A
The picture can't be
displayed.
C
A
Left facing blue arrow returns you to the dosing pattern selection screen.
B
Press the inoculum level to select the required dispensing volume.
C
Plate section.
Plate reset icon. This icon is used to reset the dosing pattern ready for a new plate. This
D
is only required if a full plate inoculation is not completed.
8. Press the center of the section you want to inoculate.
9. During inoculation, the section being inoculated will be coloured yellow until inoculation has been
completed. At the end of the inoculation process, the section will turn red.
Sensititre AIM™ Instructions for Use
The picture can't be
displayed.
B
The picture can't be
displayed.
B
The picture can't
D
16