Plate Offset button allows you to switch between Sensititre and standard microtiter plates.
Troubleshooting
Loss of power
If the "Power On" LED on the front of the instrument is off and main LCD display is also off, the system
has lost power.
1. Disconnect the main supply.
2. Check the fuse in the main fuse holder, which is located above the main socket on the
instruments back panel. (See technical specification for correct type and rating). Replace the fuse
if necessary.
Instrument errors
Problem
E1
Sensititre AIM™ Instructions for Use
Cause
NVR Checksum Error
Solution
Restart the instrument.
If the issue persists, contact Service
Engineer.