ECG recording and analysis
As soon as the electrodes are attached to the patient's
bare chest, the device automatically starts recording and
analysing the ECG. ECG recording and analysis is per-
formed continuously until the electrodes are removed
from the patient and the MEDUCORE Easy is switched off
by the lid being closed.
Defibrillation
If the ECG analysis provides an indication that defibrillation
is required (pulseless ventricular tachycardia (PVT) or ven-
tricular fibrillation (VF)), the MEDUCORE Easy prepares to
deliver a shock. The device then requests the user to initi-
ate the shock.
For all other heart rhythms the device will prompt the user
to perform CPR.
Documentation of use
The MEDUCORE Easy stores ECG and event data. These
can be read out later, for example for processing data from
individual uses. To do this, use the EasyView documenta-
tion and configuration software.
Self-test
The MEDUCORE Easy performs a self-test at regular inter-
vals and every time it is switched on. Device status is indi-
cated by LEDs on the front of the device. Self-test results
can be called up via the EasyView documentation and
configuration software.
EN
12
Description of device