9. Troubleshooting

9.1 Visual and acoustic error messages

Malfunction
In stand-by mode (lid
closed) a bleeping noise
can be heard every four
minutes.
Device does not switch on
when the lid is opened.
The message: "Device is
not ready for use." is
heard after lid is opened.
The message: "Battery is
low." is heard after lid is
opened.
Red status indicator on
(flashes during stand-by,
on continuously during
operation).
Green and yellow status
indicators on (flashing
during stand-by,
continuously on during
operation) .
Yellow status indicator on
(flashes during stand-by,
on continuously during
operation).
Caution!
We recommend having maintenance measures
like servicing and repair work carried out by the
manufacturer, Weinmann, or by specialists ex-
pressly authorized by the manufacturer.
Cause of fault
The device has found a
malfunction during one of its
monthly/daily self-tests.
Battery capacity exhausted.
Device defective.
Device defective.
Battery capacity low.
Battery capacity low, 3 more
shocks not possible.
Device is not ready for use, self-
test has detected a fault.
Battery/rechargeable battery
capacity low; 10/6 shocks no
longer possible.
Non-critical device fault has
occurred, e.g.: wrong time.
Remedy
Check which status indicators are
flashing and proceed as described in this
table for the status indicator in question.
Change battery (9.2, page 41).
Do not use device. Have repaired.
Do not use device. Have repaired.
Complete current use, then change
battery (9.2, page 41).
Change battery (9.2, page 41), then
continue use.
Complete current use. Then have device
repaired.
Complete current use, then change
battery (9.2, page 41).
Complete current use. Then read out
fault report using EasyView and remedy
fault if necessary.