6. Operation
6.1 Before use
EN
22
Operation
In this section you will find information about the follow-
ing topics.
•
"Before use" on page 22
•
"Performing resuscitation" on page 23
•
"After use" on page 31
•
"Administering use/device data using EasyView" on
page 33
Checking condition of device
The device performs a self-test daily and monthly. The sta-
tus indicators on the front of the device indicate the status
of the device. The device is ready for use when the green
status indicator flashes. Observe the following points be-
fore using the MEDUCORE Easy:
Warning!
Do not use the device if the red status indicator is
flashing. The device is not ready for use. Have the
device repaired by the manufacturer Weinmann
or an authorized dealer.
Caution!
If the green and yellow status indicators are flash-
ing, the capacity of the battery pack/rechargeable
battery pack is adequate for another 10/6 shocks.
Use the device for the current use. Only change
the battery pack/rechargeable battery pack after
use.