En Troubleshooting - Weinmann MEDUCORE Easy Descripción Del Aparato E Instrucciones

Ocultar thumbs Ver también para MEDUCORE Easy:
Idiomas disponibles

Idiomas disponibles

Malfunction
One or more status
indicators/display fields do
not come on briefly after
switching on.
Voice prompt: "Attach the
electrodes to the bare
chest!" is repeated, even
though electrodes have
been attached.
Shock cannot be delivered
despite shock key
flashing.
Voice prompt "Movement
detected. Stand clear of
the patient!" is heard.
Green status indicator
does not flash when lid
closed.
EN
40
Troubleshooting
Cause of fault
One or more status indicators/
display fields defective.
Electrode plug not plugged in
properly.
Electrodes not properly attached.
Electrodes are in contact with one
another.
Electrodes defective.
Incorrect electrodes.
Device damaged.
Device damaged.
MEDUCORE Easy detects artefact
and performs another analysis.
Battery pack/rechargeable battery
pack has discharged completely.
Status indicator defective.
Remedy
Complete current use, then
have device repaired.
Plug in plug correctly.
Press electrodes firmly onto dry and, if
necessary, shaved skin.
Check position of electrodes.
Replace electrodes, otherwise do not use
device.
Use only original Weinmann electrodes.
Have repaired.
Have repaired.
Do not touch or move patient during
analysis.
Insert new battery pack/rechargeable
battery pack.
Have repaired.