Power consumption
• BOS, inhibited: 7 µA
• BOS, 100% pacing: 18 µA
Average service time
Average service times are precalculated using the battery manufacturer's technical
specifications, a basic rate of 60 bpm in DDDR mode and the setting of different pulse
amplitudes and lead impedances. These apply to Evia HF as well as Evia HF-T.
Service times triple-chamber
For triple-chamber devices, the following times (in years) result:
Amplitude
Atrial pacing,
Impedance:
Pulse width:
Biventricular
500 ohm
0.4 ms
pacing 100%
A+RV+LV
A: 2.0 V
0%
10.6
RV: 2.2 V
30%
10.2
LV: 2.4 V
50%
9.9
100%
9.4
A: 2.5 V
0%
10.3
RV: 2.5 V
30%
9.8
LV: 2.5 V
50%
9.6
100%
9.0
A: 2.5 V
0%
8.8
RV: 2.5 V
30%
8.5
LV: 3.5 V
50%
8.3
100%
7.8
A: 3.0 V
0%
8.8
RV: 3.0 V
30%
8.3
LV: 3.0 V
50%
7.9
100%
7.3
en • English
Impedance:
Impedance:
500 ohm A+RV,
1000 ohm
800 ohm LV
A+RV+ LV
11.1
12.2
10.8
11.9
10.4
11.8
9.8
11.3
10.8
12.0
10.4
11.7
10.1
11.4
9.4
10.9
9.8
10.8
9.3
10.6
9.1
10.4
8.6
10.0
9.4
10.8
8.8
10.3
8.5
10.1
7.8
9.5
21
Amplitude
Atrial pacing,
Impedance:
Pulse width:
Biventricular
500 ohm
0.4 ms
pacing 100%
A+RV+LV
A: 3.5 V
0 %
5.3
RV: 3.5 V
30%
5.1
LV: 5.0 V
50%
4.9
100%
4.6
Shortening of the service time after long storage period
Depending on the storage period, the service time from the beginning of service BOS to
the replacement time ERI decreases as follows:
• After 1 year by three months
• After 1.5 years by four months
Impedance:
Impedance:
500 ohm A+RV,
1000 ohm
800 ohm LV
A+RV+ LV
6.4
7.7
6.1
7.3
5.8
7.2
5.3
6.8