Parameter
Range of values
Change of the basic rate
OFF
with mode switching
+5 ... (5) ... +30 bpm
Rate stabilization
OFF; ON
Refractory periods
Parameter
Range of values
Atrial refractory period
AUTO
Atrial refractory period in
300 ... (25) ... 775 ms
the modes AAI(R); AAT(R);
DDT
PVARP
AUTO
175 ... (5) ... 600 ms
PVARP after PVC
PVARP + 150 ms (max: 600 ms) is
automatically programmed
Ventricular refractory
200 ... (25) ... 500 ms
period
Blanking periods
Parameter
Range of values
Far-field protection after Vs 100 ... (10) ... 220 ms
Far-field protection after Vp 100 ... (10) ... 220 ms
Ventricular blanking period
30 ... (5) ... 70 ms
after Ap
PMT protection
Parameter
Range of values
PMT detection/termination
OFF; ON
VA criterion
250 ... (10) ... 500 ms
Pacing and Sensing: Triple-Chamber
Standard
Pulse amplitude and pulse width
+10 bpm
Parameter
OFF
Pulse amplitude A
Pulse width A
RV/LV pulse amplitude 0.2 ... (0.1) ... 6.0 ... (0.5) ... 7.5 V
Standard
RV/LV pulse width
AUTO
Sensitivity
350 ms
Parameter
Sensitivity A
AUTO
RV sensitivity
400 ms
LV sensitivity
250 ms
Atrial capture control
Parameter
Standard
Atrial capture control
100 ms
150 ms
Min. amplitude
30 ms
Threshold test start
Safety margin
Search type
Interval
Standard
Time of day
ON
350 ms
16
Range of values
0.2 ... (0.1) ... 6.0 ... (0.5) ... 7.5 V
0.1 ...(0.1) ... 0.5; 0.75; 1.0; 1.25; 1.5 ms 0.4 ms
0.1 ... (0.1) 0.5; 0.75; 1.0; 1.25; 1.5 ms
Range of values
AUTO
0.1 ... (0.1) ... 1.5 ... (0.5) ... 7.5 mV
AUTO
0.5 ... (0.5) ... 7.5 mV
OFF; AUTO
0.5 ... (0.5) ... 7.5 mV
Range of values
ATM (monitoring only)
ON; OFF
0.5 ... (0.1) ... 4.8 V
2.4; 3.0; 3.6; 4.2; 4.8 V
0.5 ... (0.1) ... 1.2 V
Interval; time of day
0.1; 0.3; 1; 3; 6; 12; 24 h
00:00 ... (10 min) ... 23:50
Standard
3.0 V
3.0 V
0.4 ms
Standard
AUTO
AUTO
AUTO
Standard
ON
1.0 V
3.0 V
1.0 V
Time of day
24 h
02:00 hh:mm