Rate adaptation via accelerometer
R modes:
Parameter
Range of values
Sensor gain
1 ... 23
Max. activity rate
80 ... (5) ... 180 bpm
Automatic gain
OFF; ON
Sensor threshold
Very low; low; medium; high; very high
Rate increase
1 ... (1) ... 10 bpm/cycle
Rate decrease
0.1; 0.2; 0.5; 1.0 bpm/cycle
Rate fading
OFF; ON
Preset programs: Triple-Chamber
Standard and safe program
Only the auto-initialization function is activated as a factory setting. All the other func-
tions of the standard program are deactivated.
Parameter
Mode (after auto-initialization: DDD)
Basic rate
Night program
Rate hysteresis
Upper rate
Dynamic AV delay
AV hysteresis
Sense compensation
AV Safety Delay
VV delay
LV T-wave protection
Far-field protection after Vs
Far-field protection after Vp
Ventricular blanking period after Ap
Standard
4
120 bpm
ON
Medium
4 bpm/cycle
0.5 bpm/cycle
OFF
Standard program Safe program
DDDR
VVI
60 bpm
70 bpm
OFF
OFF
OFF
OFF
130 bpm
Low
OFF
-45 ms
100 ms
ON
ON
100 ms
150 ms
30 ms
Parameter
PMT protection
VA criterion
Magnet response
Pulse amplitude A
Pulse amplitude RV, LV
Pulse width A
Pulse width RV, LV
Sensitivity A
Sensitivity RV, LV
Refractory period A
Refractory period RV
Refractory period LV
Mode switching
Onset criterion
Resolution criterion
Intervention rate
Switches to
Basic rate with mode switching
Rate stabilization with mode switching
2:1 lock-in protection
PVARP
PVARP after PVC
Lead configuration, automatically determined and set:
Pacing polarity A/RV/LV
Sensing polarity A/RV/LV
Automatic lead check A/RV/LV
Active capture control A/RV/LV
Atrial overdrive pacing
18
Standard program Safe program
ON
350 ms
AUTO
AUTO
3.0 V
3.0 V
4.8 V
0.4 ms
0.4 ms
1.0 ms
AUTO
AUTO
2.5 mV
AUTO
250 ms
300 ms
200 ms
200 ms
ON
5-out-of 8
5-out-of 8
160 bpm
DDIR
+10 bpm
OFF
ON
AUTO
400 ms
Unipolar
Unipolar
Unipolar
Unipolar
ON
ON
OFF
OFF