Prevena™ 125 Therapy Unit Indicators and Alerts
Caution Indicator
Audible Alarm (Muted)
Battery Level
Leak Alert
Canister Full Alert
Low Battery Alerts
System Error Alert
Device Lifecycle Expired
Visual Alerts - Solid LEDs cannot be turned off by the user. Visual alerts
will only stop when the alert condition has been corrected.
Audible Alerts - Repeated beeps (which in some cases will increase in
volume) can be temporarily muted (paused) by pressing the On / Off
button once. The audible alert will re-occur after 60 minutes unless the
alert condition has been corrected.
The Prevena™ 125 Therapy Unit will provide audible and visual alerts as
shown below.
If alert conditions cannot be corrected, patient should contact
treating physician. For further product support see Customer Contact
Information section on page 23.
One beep, one solid yellow light.
See Correcting a Leak Condition section on page 14.
When a leak condition is corrected the therapy unit will cancel alert.
There may be a delay between when the leak is corrected and when
the alert is discontinued.
Two beeps, one solid yellow light.
Visually inspect canister. If full or near full, turn therapy unit off and call
the treating physician immediately.
Low Battery Alert - One slow beep, one solid yellow light. Change
batteries within six hours.
Critical Battery Alert - One beep, repeating rapidly, increasing in
volume, one solid yellow light. Change batteries immediately.
Battery replacement will cancel the alert. See the Battery Replacement
on the Prevena™ 125 Therapy Unit section on page 16.
One beep, repeating rapidly, increasing in volume, two solid yellow
lights. Turn the unit off and then on again.
If alert continues, notify the treating physician.
Three solid yellow lights. If the lifecycle expires while therapy unit is on,
a beep will sound for 15 seconds and then automatically shut off.
If the therapy unit is off, has exceeded the run time, and an attempt is
made to turn the therapy unit on, the therapy unit will alert for three
seconds and automatically shut off. Call the treating physician.
13