PREVENA™ 125 THERAPY UNIT INDICATORS AND ALERTS
Caution Indicator
Audible Alarm (Muted)
The PREVENA™ 125 Therapy Unit will provide audible and visual alerts as shown below.
Battery Level
Leak Alert
Canister Full Alert
Low Battery Alerts
System Error Alert
Device Lifecycle Expired
Visual Alerts - Solid LEDs cannot be turned off by the user. Visual
alerts will only stop when the alert condition has been corrected.
Audible Alerts - Repeated beeps (which in some cases will increase in
volume) can be temporarily muted (paused) by pressing the On/Off
button once. The audible alert will re-occur unless the alert condition
has been corrected.
If alert conditions cannot be corrected, patient should contact
treating physician. For further product support see Customer Contact
Information section.
Audible Alert can be muted for eight hours.
One beep, one solid yellow light.
See Correcting a Leak Condition section.
When a leak condition is corrected the therapy unit will cancel alert.
There may be a delay between when the leak is corrected and when
the alert is discontinued.
Audible Alert can be muted for one hour.
Two beeps, one solid yellow light.
Visually inspect canister. If full or near full, turn therapy unit off and call
the treating physician immediately.
Audible Alert can be muted for one hour.
LOW - One slow beep, one solid yellow light. Be prepared to change
batteries.
CRITICAL - One beep, repeating rapidly, increasing in volume, one
solid yellow light. Change batteries immediately.
Battery replacement will cancel the alert. See Battery Replacement
on the PREVENA™ 125 Therapy Unit section.
Audible Alert can be muted for one hour.
One beep, repeating rapidly, increasing in volume, two solid yellow
lights. Turn the unit off and then on again.
If alert continues, notify the treating physician.
Three solid yellow lights. If the lifecycle expires while therapy unit is on,
a beep will sound for 15 seconds and then automatically shut off.
If the therapy unit is off, has exceeded the run time, and an attempt is
made to turn the therapy unit on, the therapy unit will alert for three
seconds and automatically shut off. Call the treating physician.
15