Air Line Requirements And Recommendations; Electrical Requirements; Unpacking The System; Power Cord/Power Connection - DENTSPLY Cavitron Instrucciones Para El Uso

Jet-escarificador ultrasonico y unidad profilactico de pulido neumatico
Ocultar thumbs Ver también para Cavitron:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

water supply system, the dental office water line should
be thoroughly flushed prior to connection to the System.
5.3 Air Line Requirements &
Recommendations
• An air supply line with a user-replaceable filter assembly
is factory installed on your Cavitron JET Combination
system. Do not disconnect the air supply line. A filter-
mounting bracket is included for hanging the air filter.
The clear bowl should hang downward allowing for mois-
ture separation and drainage of water from the air filter.
See 9.0 System Care for replacement instructions.
• Incoming air supply line pressure to the System must
be 65 psig (448kPa) to 100 psig (690kPa). If your office
air line pressure is above 100 psig (690kPa), install an
air pressure regulator on the supply line to your
Cavitron
JET Combination System.
• A manual shut-off valve on the office air supply line
should be used so that the air line can be completely
shut-off, and the line pressure relieved when the office
is unoccupied.
• Your System must be supplied with clean, dry air to help
prevent water condensation from forming in the air sup-
ply line which may cause the system to malfunction. In
addition to the air filter supplied with your system, it is
strongly recommended that an air dryer be used on the
compressor line supplying the System.

5.4 Electrical Requirements

Refer to Section 11: Specifications.

5.5 Unpacking the System

Carefully unpack your Cavitron
verify that all components and accessories are included:
1. Cavitron
JET Combination System with water sup-
ply line, air supply line, handpiece and cable connector
assembly factory installed.
2. Air filter assembly.
3. Detachable AC Power Cord set.
®
4. Cavitron
JET Airpolishing Insert with cleaning tool.
5. Dual position Foot Control Assembly.
®
6. Cavitron
Ultrasonic Inserts (quantity optional).
7. Efficiency Indicator for Cavitron
8. JetShield
Aerosol Reduction Device.
9. Literature Packet.
®
10. PROPHY-JET
Prophy Powder.
11. JET-Fresh
Prophy Powder.
JET Combination System and
®
Inserts.
Section 5: Installation Instructions con'd

5.6 Power Cord/Power Connection

• Verify the Power Control ON/OFF rocker switch located
at the left front underside of the System is set to the OFF
position before proceeding.
• Plug the detachable AC cord into the back of the System.
• Plug the 3 prong grounded plug into an approved outlet.
ON/OFF rocker switch
5.7 Foot Control Cable Assembly
Connection
Align the Foot Control plug with the receptacle on the back
of the System and push it in until firmly seated.
ENGLISH
5