4.1 Set the Drive Air Pressure

Supply air, and adjust to 0.3 MPa by pulling the regulator
knob upward and turning. When adjustment is made,
push the knob down to lock.

4.2 Set the Water Supply

For use with water supplied, set the water ON/OFF switch
to the ON side. Use of Water Supply Bottle: Push the
water mode select switch until it stops (Fig.11). Use of
City Water: Pull the water mode select switch until it stops
(Fig. 11). Open the main cock of the tap water.
• Properly set the water mode select switch
because if it is stopped at an intermediate
position, air is released structurally.

4.3 Operation

Stepping on the foot pedal starts rotation. If the water
ON/OFF switch is set to the ON side, water supply is also
started.