9.3 Recommended separation distances

Recommended separation distances between portable and mobile RF communications equipment
and the ARTROMOT®-E2/-E2 compact
The ARTROMOT®-E2/-E2 compact is intended for use in an electromagnetic environment in which radi-
ated RF disturbances are controlled. The customer or the user of the ARTROMOT®-E2/-E2 compact can
help prevent electromagnetic interferences by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the ARTROMOT®-E2/-E2 compact as recom-
mended below, according to the maximum output power of the communications equipment.
rated maximum output
power of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation dis-
tance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter,
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1: For calculation of the recommended separation distance of transmitters in the frequency range
from 80 MHz to 2.5 GHz an additional factor of 10/3 was taken into account to reduce the probability of
mobile/portable communications equipment brought into the patient environment by accident causing
any malfunction.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.

10. Contact

We would be happy to answer any ques-
tions you may have about our products
and services.
n ORMED.DJO international
Please contact your local dealer or the
ORMED.DJO headquarters in Germany.
n Headquarters Germany
ORMED.DJO
ORMED GmbH & Co. KG
Merzhauser Strasse 112
D-79100 Freiburg - Germany
Tel. +49 761 45 66-01
Fax +49 761 45 66 55-01
n Internet
www.ormed-djo.de
separation distance according to frequency of transmitter
150 kHz to 80 MHz
öä
d = 1.2
P
0.12
0.12
0.38
0.38
1.2
1.2
3.8
3.8
12
12
Warranty:
2 years (mechanical parts)
2 years (electronics)
Sales:
ORMED.DJO
ORMED GmbH & Co. KG
Merzhauser Strasse 112
D-79100 Freiburg - Germany
87
m
80 MHz to 800 MHz
800 MHz to 2.5 GHz
öä
d = 1.2
P
0.23
0.73
2.3
7.3
23
öä
d = 1.2
P
Tabla de contenido
loading

Este manual también es adecuado para:

Artromot-e2 compact

Tabla de contenido