• To modify the value, press the "+" (plus)
or "-" (minus) key.
Select 25%, for instance.
• Save the programmed values by press-
ing the STOP key, then press the START
key to begin the treatment.
This is what happens:
Extension/flexion is treated three times first
(motor A). Motor B is positioned at 25%
of the programmed range of motion for
pronation and supination.
Subsequently, pronation/supination is
treated once (motor B). Motor A is posi-
tioned at 75% of the programmed range of
motion for extension and flexion.
Note!
The percentage values can be modi-
fied only when the special function is
enabled (check mark).
n Total therapy time
Under menu item "total therapy time"
you can view each patient's total therapy
time (duration of all the patient's therapy
sessions).
Deleting the stored therapy time:
Press and hold the parameter key for
5 seconds or select the "new patient"
function.
n Extension/flexion therapy documen-
tation
This is a special function of the
ARTROMOT®-E2/-E2 compact which
allows the entire therapy documentation
to be represented.
The carriage run times as well as the
range of motion of the sessions are
recorded.
The collected data are presented graphi-
cally in the form of a coordinate system
(X-axis = range of motion / Y-axis = time)
where the upper curve illustrates the
trend of the flexion movement and the
lower curve the trend of the extension
movement.
n Pronation/supination therapy docu-
mentation
This is a special function of the
ARTROMOT®-E2/-E2 compact which
allows the entire therapy documentation
to be represented.
The carriage run times as well as the
range of motion of the sessions are
recorded.
The collected data are presented graphi-
cally in the form of a coordinate system
(X-axis = range of motion / Y-axis = time)
where the upper curve illustrates the
trend of the supination movement and
the lower curve the trend of the prona-
tion movement.
71