In normal mode the display indicates the
symbol "OFF" during therapy, instead
of the programmed extension/flexion
angles.
Default: motor A ON
n Motor B ON/OFF
To permit fully isolated movements, the
motors can individually be turned on
and off. Motor A effects the extension
and flexion movements, while motor
B effects the pronation and supination
movements.
For an isolated extension/flexion move-
ment, set motor B to the appropriate
position (pronation/supination), then
turn the motor off.
In normal mode the display indicates the
symbol "OFF" during therapy, instead of
the programmed pronation/supination
angles.
Default: motor B ON
Note!
Please note that one motor (A or B) must
always be turned on. Otherwise the
following symbol will be displayed when
you press START:
LEVEL 4:
n Isolation protocol
For this special function, both motors
are turned on, but they will never move
simultaneously.
These are the steps of the function:
• To begin with, motor A moves to the
programmed maximum values for
extension and flexion for three cycles
and then stops. In this phase, motor B is
disabled (symbol for motor B: OFF)
• Then motor B will move to the pro-
grammed maximum values for prona-
tion and supination for one cycle and
stops, then the full cycle will start over
with motor A. Motor A is disabled while
motor B is operating (symbol for motor
A: OFF)
• The stop position at the end of the last
cycle (for both extension/flexion and
pronation/supination) can be selected
in steps of 25% between 0% and 100%
of the programmed maximum range of
motion.
• Steps 1 and 2 can be repeated as often
as needed. You can stop treatment
with the STOP key; after the pro-
grammed therapy duration, it will stop
automatically.
During warm-up, the display will show
the symbol
hand corner.
Default: disabled
Follow these steps to program the special
function:
• Access menu level 4, then press the
parameter key to select the special func-
tion
.
The display will indicate:
- the symbol representing the special
function
- the status of the special function (circle
with or without check mark)
- two boxes with the current percentage
of the stop positions for motor A and B
- a check mark in a circle.
• You activate the function with the "+"
key and deactivate it with the "-" key.
In this case it is not possible to activate/
deactive the function by pressing the
parameter key again.
A check mark in the circle identifies the
activated function.
• Next press the parameter key of the
isolation protocol. The marking moves
to the upper percentage value, allowing
you to enter the stop position for
motor A.
• To modify the value, press the "+" (plus)
or "-" (minus) key.
Select 75%, for instance.
• Press the parameter key again when
done. The marking automatically moves
to the lower box, allowing you to adjust
the stop position of motor B.
70