7. Environmental Protec-
tion Statement
The product described in this operation
manual must not be disposed of with
unsorted household or municipal waste. It
requires separate disposal. Please contact
ORMED.DJO for information about the pos-
sible recycling of the product.
8. Specifications
Input ratings:
Rated current - motor:
Power consumption
Fuses:
Protection class:
Applied part:
Dimensions (transport):
ARTROMOT®-E2
Length:
Width:
Height:
ARTROMOT®-E2 compact
Length:
Width:
Height:
Adjustment ranges (min./max.):
Height adjustment:
(ARTROMOT®-E2 measured from the seat)
Forearm length:
Seat height:
ARTROMOT®-E2
Weight:
ARTROMOT®-E2
ARTROMOT®-E2 compact
Materials used:
MDD:
Standards compliance: IEC 60601-1:1990
100 – 240 V AC /
50 – 60 Hz
2 A max.
33 VA
2 x T1A
I
type B
87.5 cm
57.5 cm
58 cm
87.5 cm
57.5 cm
29 cm
35 – 71 cm
29 – 46 cm
48 cm
25 kg
17 kg
ABS, POM (Delrin
100), PUR, PA,
FR4, aluminum,
stainless steel,
brass
class 2a
+ A1:1993
+ A2: 1995
No. 601.1-M90
UL 2601-1
83
EMC
(electromagnetic
compatibility)
Ambient conditions (storage, transport)
Ambient temperature: -24 ºC to +60ºC
Relative humidity:
Atmospheric pressure:
Ambient conditions (operation)
Ambient temperature: +10 ºC to +40ºC
Relative humidity:
Atmospheric pressure:
____________________________________
Subject to change without notice (06/2008)
IEC 60601-1-
2:2001
20 % to 85 %
700 hPa to
1060 hPa
30 % to 75 %
700 hPa to 1060
hPa