Fuse replacement
m Warning!
Patient hazard, equipment malfunction
and damage—
The replacement of fuses must be
referred to specialists as defined in IEC
60364 or other applicable standards
(e.g. biomedical technicians, electricians,
electronics installers).
Fuses used must be T1A fuses.
• Before replacing fuses turn off the
ARTROMOT®-E2/-E2 compact and
remove the power cord from the wall
outlet.
• Use an appropriate tool to remove the
fuse holder (20/8) situated between the
power switch (19) and the power con-
nector (18) (Fig. 1).
• Replace the fuses and reinsert the fuse
holder (Fig. 2). Check that the fuse
holder locks properly into place.
Fig. 1
Fig. 2

6.3 Transport

6.3.1 ARTROMOT®-E2
The following operating steps must
be completed before transporting the
ARTROMOT®-E2:
1. Activate the "transport setting" func-
tion
in the menu (see also 5.3)
and start the ARTROMOT®-E2.
2. Turn off the ARTROMOT®-E2 power
switch (19).
3. Remove the power cord (18) and dis-
connect the motion element (21) and
the programming unit (32).
4. Open locking screw (4) and pull out
the armrest (14) for the healthy arm.
5. Grasping the motion element at the
bracket (30) for adjustment of the posi-
tioning angle, open the locking screw
(4) and pull out the motion element.
6. Set horizontal adduction/abduction to
0° (indexing knob 1).
7. Open the locking screw for adjustment
of the backrest (10), fold the backrest
to the front all the way until flat on the
seat and tighten the screw.
8. Remove the two locking pins (22). Pull
out the legs towards the sides, reverse
and reinsert. Then put the locking pins
back in place.
9. Only use the original shipping box
for transporting the device. Ormed
GmbH & Co. KG cannot be held liable
for transport damage, if the device was
not shipped in its original shipping
box.