5.2 Programming the
ARTROMOT
-E2/
®
-E2 compact
Different programming levels are provided
to program the
ARTROMOT
-E2/-E2 compact.
®
You change between levels by pressing the
MENU key repeatedly.
The currently selected level is indicated on
the display.
The following treatment values, settings
and information can be entered/viewed
on the programming unit (15):
LEVEL 1:
- extension
- flexion
- pronation
- supination
LEVEL 2:
- pause
- therapy timer
- speed
- warm-up protocol
LEVEL 3:
- reverse on load - motor A
- reverse on load - motor B
- motor A ON/OFF
- motor B ON/OFF
LEVEL 4:
- isolation protocol
- total therapy time
- therapy documentation:
extension/flexion
- therapy documentation:
pronation/supination
LEVEL 5:
- stretching in extension
- stretching in flexion
- stretching in pronation
- stretching in supination
LEVEL 6:
- EROM repeat
extension/ pronation
- EROM repeat
flexion/supination
LEVEL 7:
- transport setting
- new patient
- operating mode
synchronized/
non-synchronized
- Service menu
Note!
• While you adjust the values, the
carriage will move to the set range.
This allows you to easily and quickly
determine the ROM where the patient
does not experience any pain.
• The angle entered last for the respec-
tive directions of motion will be saved
in each case.