LEVEL 2:
n Pauses
Pauses occur at each programmed maxi-
mum value.
These are the two pause points:
− maximum extension value and maxi-
mum pronation value
− maximum flexion value and maxi-
mum supination value.
Pauses are separately adjustable for
motor A and motor B between 1 second
and 60 minutes. Pauses can be set in
steps of 1 second to any value between
0 and 59 seconds, and in steps of 1
minute to any value between 1 and 60
minutes.
Default: no pause
Follow these steps to program pauses:
• Access menu level 2, then press the
parameter key to select the special
function
.
• The display shows the symbol for the
special function and the selected pause
durations.
The display shows the symbol for the
special function and the selected pause
durations.
top: flexion/supination pause
bottom: extension/pronation pause
.
The flexion/supination pause duration
appears in a box.
• To change the flexion/supination pause,
press the "+" or "-" key.
Select a duration of 10 seconds, for
instance.
• Press the parameter key again when
done. The frame automatically moves to
the line below, indicating that the exten-
sion/pronation pause can be adjusted
.
The parameter key is only used to
toggle between these two functions.
• To modify the value, press the "+" (plus)
or "-" (minus) key.
Select a duration of 5 seconds, for
instance.
• Save the programmed values by press-
ing the STOP key, then press the START
key to initiate the treatment.
.