Operating Manual X-zone Com - Sending Text Messages to X-Zone Com
7.1.4 Query for a group's maximum measured values
X-zone Com transmits a list that contains the maximum measure values of each gas monitored by the
group, including types of alarms present and device IDs. In the instance of oxygen, the reading that
appears in the list is the minimum value.
[PIN] max
Text Message Reply Contents
X-zone Com device name
Gas type
Concentration
Unit of measure
Alarm type
Identification number of the X-zone
5x00 with the maximum value in the
group
Example: Maximum value
Fig. 7.6: Message displayed on a smart phone following a query to obtain the maximum values of all gases moni-
tored in a group. In this example, the group device with ID 1 read the ch4 value. The master device with 0 ID
measured the other gasses. For O
Fig. 7.7: Smart phone display if the connected X-zone 5x00 is not a group device. For such instances, use the
"local" command.
Value Ranges
User-specified ASCII character succession
Abbreviated gas name of X-am 5x00
0 to 10,000
Vol%, %UEG, %LEL, %LIE, ppm, mg/m3,
ppb, mbar, hPa, kPa, mg/L
A1, A2
0 to 15
and CO
there is an A1, respectively an A2 Alarm.
2
2