Coltene CanalPro CL Instrucciones Para El Uso página 23

Ocultar thumbs Ver también para CanalPro CL:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 38
7. Cleaning
(1) Lubricating Contra Angle Head
- Lubricate the contra angle head only.
1. Lubricate it before each autoclaving or once a day if
you do not autoclave. Screw the F type spray nozzle
onto the spray nozzle by approx. 10 turns.
2. Insert the F type spray nozzle into the rear part of the
contra angle head and lubricate the head for 2-3 sec-
onds. If the contra angle head does not push into
nozzle enough, oil may not go around into the hand-
piece and it may flows backward.
(2) Cleaning Motor Handpiece
When the motor handpiece becomes dirty, wipe it off with a cotton cloth moistened with rubbing alco-
hol.
(3) Cleaning Charger
When the charger becomes dirty, wipe it off with a cotton cloth moistened with surgical solution.
CAUTION
- Do not lubricate the motor handpiece.
- Before mounting the lubricated contra angle head to the motor handpiece, wipe off excess oil. Stand it
on its end or lean it in the proper position for gravity draining. Mount it after excess oil has been com-
pletely drained.
- Hold the contra angle head securely to prevent it from flying off by the pressure of the spray.
- Hold the spray can upright.
- When cleaning the motor handpiece, do not use any solvent such as benzene and, thinner.
8. Sterilization
Sterilize the contra angle head only.
For the sterilization method, we recommend the autoclaving. Sterilization is required first time you use and
after each patient as noted below.
„ Autoclaving
1. Brush the dirt off the surface of the contra angle head, and wipe it off with a cotton cloth moistened with
surgical spirit. Do not use a metal brush.
2. Lubricate the head (Refer to the "7 (1) Lubricating Contra Angle Head. " )
3. Insert the head into an autoclave pouch and seal it.
4. Sterilization according to the current regulations at maximum 135°C (275°F).
5. Keep the handpiece in the autoclave pouch to keep it clean until you use it.
CAUTION
- Do not heat sterilize the motor handpiece.
- Do not wipe with, or clean or immerse in, high acid water or sterilizing solutions.
44
F type spray nozzle
Contra angle head
Fig. 11
9. Motor Cap
When the contra angle head is removed from the motor
handpiece for battery charging, lubrication, or steriliza-
tion, mount the motor cap onto the motor handpiece
to prevent debris from entering.
10. Safety System
The motor handpiece monitors temperature of the batteries. If the batteries can potentially become abnor-
mally hot, this system functions and the motor handpiece automatically stops. In such a case, wait until the
motor handpiece cools down sufficiently. If this safety system functions repeatedly, either the batteries or
motor handpiece is not in the normal operating condition and you should contact your dealer.
11. Changing Batteries
The motor handpiece uses rechargeable batteries. They can be recharged 300-500 times, depending on
the use conditions of the motor handpiece. If the operating time becomes shorter or the rotation speed
becomes slower, although the "MEMORY EFFECT" described in "
may be at the end of their life expectancy. In such a case, ask your dealer to replace the batteries or replace
them with new ones yourself. (Refer to the " Changing Batteries. " ) When replacing them by yourself, be sure
to observe the following "
CAUTIONS ON CHANGING BATTERIES. " Please note that Coltène/Whaledent
shall not be liable for any malfunction or failure resulting from you changing the batteries yourself and not
following the "
CAUTIONS ON CHANGING BATTERIES. "
CAUTIONS ON CHANGING BATTERIES
- Do not open any part other than the battery cover.
- Use only batteries as specified by Coltène/Whaledent.
- Designated Batteries : AAA (marking may be different) nickel metal hydride batteries, nominal 1.2 V
- Do not use non-charging type batteries such as alkaline batteries and manganese batteries. Charging
with these batteries may cause fluid leaks, explosion or chlorine gas generation.
- Should the battery fluid leak and get into your eyes, immediately wash thoroughly with clean water and
seek medical attention.
- Should the battery fluid leak and adhere to skin or clothing, immediately wash the exposed skin thor-
oughly with clean water and completely wash away the fluid. Failure to do so may result in a skin irrita-
tion.
- Always replace two batteries of the same type by the same manufacturer at the same time. Using batter-
ies of different-types, an exhausted battery with a fully charged one, or a new battery with an old one
may cause a fluid leak or explosion.
- Do not work with wet hands. Failure to do so may result in rust formation on battery terminals or mois-
ture intrusion inside, and could cause failure of the product.
- Do not reverse positive (+) and negative (-) when fitting the batteries.
Motor Cap
Motor Handpiece
Fig. 12
NOTICE" is not applicable, the batteries