18
Patient-side
BIOTRONIK cable
connection
Recommended connections
Screw terminals
PK-83B
Screw terminals
PK-83
Possible connections
Screw terminals
PK-175
Alligator clips
PK-141
Alligator clips
PK-67-S/L with PA-4
Alligator clips
PK-67-S/L with PK-155
Alligator clips
PK-82
Connection
BIOTRONIK cable
patient-side
Recommended connections
IS-1 connector
PK-67-S/L with PA-2
port
Possible connections
Alligator clips
PK-141
Alligator clips
PK-67-S/L with PA-4
Alligator clips
PK-67-S/L with PK-155
Alligator clips
PK-82
Note:
Heart wire with break-off needle or with flexible end
(max. 2.3 mm in diameter)
Implanted lead with IS-1 connector
For dual-chamber cables (PK-141, PK-175, PK-67-S/L),
Reocor S uses only the ventricular channel!
Polarity
Reocor S principally paces in bipolar mode, but it can be used
with bipolar or unipolar temporary pacing leads.
If unipolar leads are used, two leads must be connected.
Separating connections
Disconnect patient cables from the temporary pacing leads of
the patient or disengage the direct connection.
Separating Redel plug
•
Retract the retaining ring at the Redel plug and pull the
Redel plug off the Redel port.
Device-side
Reocor S connection
connection
Redel plug
Redel adapter
2-mm plug
2-mm connector port
Redel plug
Redel adapter
Redel plug
Redel adapter
Redel plug
Redel adapter
Redel plug
Redel adapter
2-mm plug
2-mm connector port
Connection
Reocor S
device-side
connection
Redel plug
Redel adapter
Redel plug
Redel adapter
Redel plug
Redel adapter
Redel plug
Redel adapter
2-mm plug
2-mm connector port