Changes not
permitted
Replacement parts
and accessories
Devices
on hand
Behavior
before use
Lead connection
Reocor S must not be used in areas with a danger of
explosion.
Only the manufacturer or a party expressly authorized by
BIOTRONIK may perform corrective maintenance,
enhancements or modifications to the device.
To ensure safety compliance, use only original replace-
ment parts and accessories authorized by BIOTRONIK.
Using any other parts voids the manufacturer's liability
for any consequences, guarantee and warranty.
In case of pacemaker dependency of the patient, an
emergency pacemaker should be kept on hand.
Keep an external defibrillator, oxygen, intubation equip-
ment and emergency drugs on hand.
Before use, Reocor S should be visually inspected for
damages and dirt.
Never use a device that is damaged or shows abnormal
behavior. Replace any cable that shows even slight
damage.
Before using Reocor S, the patient cable or leads, the
user should touch the patient to equalize electrical
potentials.
It is strongly recommended to examine all set para-
meters before the leads are connected to Reocor S.
Even though Reocor S is protected from dripping water,
the device and all plugs should be kept clean and dry.
Reocor S cannot be sterilized.
The connections of Reocor S and the temporary pacing
leads must be secured and checked regularly.
The patient cable must first be connected to Reocor S
and then to the leads.
The temporary leads, to which the Reocor S is connected,
represent a low-impedance conductor to the myocar-
dium for electric current. Therefore line-powered devices
that are operated in the patient's vicinity must be
grounded in accordance with established guidelines.
When handling already implanted leads, their connector
pins and metal contact surfaces must not touch or come
into contact with electrically conductive or wet surfaces.
If the cable is disconnected from the Reocor S, it must be
reconnected immediately and the security of the connec-
tion has to be examined.
When using unipolar leads, two unipolar leads must be
used for effective pacing.