16
Figure 13: Adapters for the patient cables PK-67-L and PK-67-S
WARNING!
WARNING!
WARNING!
Caution!
Note:
•
Adapter
Figure 13 shows adapters for connection of temporary leads
to Reocor S through the dual-chamber patient cable PK-67-
L/S. The leads should be connected with the ventricular con-
nection of the adapter (labeled V).
PA-1-B and PA-1-C for the connection of touch-proof 2-mm
plugs or MHW adapters (adapters for heart wires)
PA-2 IS-1
PA-4 with alligator clips
Connection
Danger to patient by damaged cables.
Damaged cables are limited in functionality and pose a
danger to patients. Do not use damaged cables.
Danger from loss of function.
Damp cables have limited functionality and pose a danger to
patients. Do not use damp cables.
Danger from electrical currents.
Unused cable contacts can conduct electrical currents to
patients. Adhere unused cable contacts close to the patient.
Allergic reactions and inflammations.
Prevent the cable from coming into contact with the patient's
wounds or skin.
Ensure correct fitting of the insulators prior to using the
cables.