8
Warning
Acoustic signal for
two seconds
Fast sequence of sounds
Acoustic signal and flash-
ing of the Pace and Sense
LEDs
Low battery LED flashes.
a)
When using the battery type MN 1604 Duracell
— Turn off all electric devices in the vicinity of Reocor S
if they can cause electromagnetic interference and
their operation is not absolutely necessary.
— Move the interference source to a location where the
interference cannot have an affect on the Reocor S.
— If safe to do: Switch Reocor S off and on again to
reset the pacemaker to interference-free operation.
— If the technical failure persists, please contact
BIOTRONIK.
Visual and Acoustic Signals
•
During the self-test after switching on Reocor S, all LEDs
light up and brief acoustic signals can be heard. The self-
test is finished after a few seconds.
•
If the self-test does not find any errors, the LEDs and
warning signals turn off.
•
When the self-test finds a defect, all LEDs flash continu-
ously and warning signals sound.
•
A required battery replacement is indicated by the flash-
ing red Low battery LED.
•
The Sense (green) LED signals sensing of a P wave or
R wave.
•
The Pace (yellow) LED signals pulse delivery.
•
The LEDs and acoustic signals also provide the following
warnings during operation:
Meaning
A pulse amplitude of < 1 V or a rate of
> 180 ppm is programmed.
Impedance outside of the permissi-
ble range
High rate protection has been
triggered; self-test failed.
ERI has been reached.
Error correction
Check whether the set values
are suitable for the patient.
Check whether all connectors
are securely plugged in.
Check whether the leads have
the desired position.
Turn the device off and return
it to BIOTRONIK.
Replace the battery; and about
36 hours
remain.
®
®
, Procell
a)
of service time