Connection Variants - BIOTRONIK Reocor S Manual Tecnico

Marcapasos externo
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 122
17
Note:
Note:
Patient-side
BIOTRONIK cable
connection
Recommended connections
Direct connection (without BIOTRONIK cables)
2 mm
PK-67-S/L with PA-1-C
Screw terminals
PK-83B with TC Adapt
Screw terminals
PK-83 with TC Adapt
Possible connections
2 mm
PK-67-S/L with PA-1-B
Alligator clips
PK-141
Alligator clips
PK-67-S/L with PA-4
Alligator clips
PK-67-S/L with PK-155
Alligator clips
PK-82
For use of cables or adapters for dual-chamber applications,
the leads must be connected to the ventricular channel
(labeled Ventricle or V).
Do not connect the patient cable to the temporary pacing lead
of the patient before the connection has been established to
the Reocor S.
Direct connection
If Reocor S is used without the Redel adapter, temporary
catheters and heart wires can be connected to the patient
cables PK-82 and PK-83 directly at the connector ports
INDIFF.+ and DIFF.-.
Patient cable
The patient cable is connected via a Redel adapter to the
Reocor S.
Fit the Redel adapter to Reocor S.
Screw the adapter in tight.
Insert the Redel plug of the patient cable into the Redel
port of the adapter.

Connection Variants

Temporary catheters with 2-mm plugs or heart wire with
2-mm adapter
You have the option to connect Reocor S directly to a tempo-
rary catheter with touch-proof 2-mm plug or to a heart wire
with 2-mm adapter, without any other cables or adapters. All
other connection options are listed in the following table:
Device-side
Reocor S connection
connection
2-mm connector port
Redel plug
Redel adapter
Redel plug
2-mm connector port
2-mm plug
2-mm connector port
Redel plug
Redel adapter
Redel plug
Redel adapter
Redel plug
Redel adapter
Redel plug
Redel adapter
2-mm plug
2-mm connector port
Tabla de contenido
loading

Tabla de contenido