Start Up

The operation of Reocor S is identical for all operating
modes. The operating steps should be carried out in the fol-
lowing order (the numbers in parentheses refer to Figure 1 on
page 10).
Insert battery.
Push protective cover down.
Prepare patient: Place the leads but do not connect them
to the pacemaker yet.
Prepare Reocor S:
Set the pacing rate with the Rate control dial (12).
Set the pacing amplitude with the Ampl. control dial (6).
Select the pacing mode with the dial Mode (13). The
device will be activated at the same time.
After the internal self-test has been completed success-
fully, the LEDs on the operating panel simultaneously
flash twice.
If the Low battery LED (3) flashes, the battery needs to be
replaced (for battery exchange, see page 20).
Connect leads; the yellow Pace LED (4) flashes in syn-
chrony with the pacing pulses.
Set the sensitivity with the Sens. control dial (7) so that
the green Sense LED (5) flashes in synchrony with every
sensed event.
A sufficient safety margin should be considered to ensure
reliable sensing.
Monitor the ECG of the patient and adjust amplitude
and/or sensitivity, if necessary.
Caution!
During use of Reocor S, the heart rate of the patient must be
monitored with an ECG monitor with alarm function.

Attachment

Reocor S must be operated either horizontally on a non-slip
surface or affixed to the patient with an armband, or from a
hanging position on the infusion stand using the hanger on
the back of the device.
To attach Reocor S to an infusion stand, unscrew the hanger
from the back of the device. This ensures safe operation and
unburdens the patient cables.