Corghi EM9780 plus Manual De Uso página 83

Idiomas disponibles
  • ES

Idiomas disponibles

6.5. Ultrasonic width sensor calibration
It is used to calibrate the ultrasonic sensor posi-
tioned on the wheel protection tube (width). It
must be executed when the machine requests
it by displaying the E4 message, or when you
notice a difference between the detected circle
width and the actual circle width.
-
select the icon
libration of the width within the configuration
programs;
-
Fix the calibration template in correspon-
dence of the threaded hole on the bell of the
oscillating unit using the screw M8 (A, Fig.24)
provided with the ultrasonic sensor;
-
use the sleeve with the wheel spacer to bring
the smoke contact with the bell of the oscillating
unit (fig.24);
-
press the LIVE button on the weighing cover
or the ENTER button to confirm the mounting
of the template;
-
slowly rotate the template towards the operator
until the parking brake is automatically enabled
(fig.25);
EM9780 plus - EM9780 C plus AutomaticLine - Operator's manual
Ultrasonic sensor ca-
24
-
The machine automatically calibrates the front
laser.
-
when the front laser is resting, slowly rotate
the template in the opposite direction to the
operator until the parking brake is automatically
enabled (fig.26);
-
Slowly lower the wheel guard (fig. 27), the ma-
chine will automatically calibrate the sensor.
If the calibration is successful, a consensus
message is displayed.
The display of the A20 message indicates
instead that:
-
the position of the calibration template during
calibration is incorrect. Place the template in
a correct way, ie by checking that the hole on
the ultrasonic sensor holder is in line with the
calibration template (Fig. 27) and repeat the
procedure.
25
UK
26
83
loading

Este manual también es adecuado para:

Em9780 c plus automaticline