DENTSPLY Cavitron SPS Instrucciones De Uso página 9

Ocultar thumbs Ver también para Cavitron SPS:
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 33
Section 5: Infection Control
5.1 General Information
As with all dental procedures, use universal
precautions (i.e., wear face mask, eyewear, or
face shield, gloves, and protective gown).
• For operator and patient safety, carefully
practice the infection control procedures detailed
in the Infection Control Information Booklet
accompanying your System. Additional booklets
can be obtained by calling Customer Service at
1-800-989-8826, Monday through Friday, 8:00
A.M. to 5:00 P.M. (Eastern Time). For areas
outside the U.S., contact your local DENTSPLY
Professional representative.
• As with high speed handpieces and other
dental devices, the combination of water and
ultrasonic vibration from the Cavitron SPS Self
Install Scaler will create aerosols. Following the
procedural guidelines in Section 9 of this manual
can effectively control and minimize aerosol
dispersion.
5.2 Water Supply Recommendations
• It is highly recommended that all dental water
supply systems conform to applicable CDC
(Centers for Disease Control and Prevention)
and ADA (American Dental Association)
standards, and that all recommendations be
followed in terms of flushing, and general
infection control procedures. (See Sections
6.2 and 10) Knowledge of and compliance
with agency guidelines, standards and
recommendations is the sole responsibility
of the Dental Healthcare Professional.
• As a medical device, Cavitron
be installed in accordance with local or national
regulations, including guidelines for water quality
(e.g. drinking water). As an open water system,
such regulation may require your Cavitron
product to be connected to a centralized water
control device that prevents water containing
contaminates from back-flow into the water
supply.
Section 6: Installation Instructions
6.1 Installation
S ee separate installation guide provided with your
system.
6.2 Water Line Requirements
• Incoming water supply line pressure to the
ultrasonic scaler must be 20 psi (137kPa)
minimum to 40 psi (275 kPa) maximum. If your
dental water system's supply line pressure is
above 60 psi, install a water pressure regulator
on the water supply line to your Ultrasonic
Scaler.
• Incoming water temperature to the Cavitron
System should not exceed 25˚C (77˚F).
If needed, a device should be installed to
®
products need to
maintain a temperature within this specification,
or a Cavitron DualSelect Dispensing System
attached to allow this system to be operated as a
closed water system.
• A manual shut-off valve on the dental water
system supply line should be used so that the
water can be completely shut-off when the office
is unoccupied.
• A filter in the dental water system supply line
is recommended, so that the particles in the
water supply will be trapped before reaching the
ultrasonic scaler.
• After the above installations are complete on
the dental water supply system, the dental office
water line should be thoroughly flushed prior to
connection to the ultrasonic system.
• After flushing system, verify there are no leaks.
6. Air Line Requirements
• Maximum drive air pressure 100 psi (689 kPa).
6. Electrical Requirements
• Power to the Cavitron SPS Self Install Ultrasonic
Scaler is supplied by the medically approved
Power Supply (GlobTek Inc. Model # GT-500160-
30) provided with your system. (Alternate Power
Supply should not be used with this system).
• Incoming power to the power supply must be 100
volts AC to 240 volts AC, single phase 50/60 Hz
capable of supplying 1.0 amps.
6.5 Unpacking the System
Carefully unpack your Cavitron SPS Self Install
System and verify that all components and
accessories are included:
®
1. Cavitron
SPS™ Self Install Scaler with
handpiece cable assembly with swivel
2. Detachable AC Power Cord
3. Medically approved Power Supply
®
4. Steri-Mate
Detachable Sterilizable Handpiece
®
5. Cavitron
Ultrasonic Inserts (optional)
6. Efficiency Indicator for Cavitron Inserts
7. Literature Packet
8. Handpiece Cable Collar
9. Velcro One-Wrap strap (2)
10. Adjustable adhesive backed Mounting Clips (3)
11. Alcohol Pad
12. 3M Dual Lock adhesive backed strips (2)
ENGLISH • 
loading