Section 7: Cavitron
Ultrasonic Scaler Description
7.1 System Controls
®
The Cavitron
SPS™ module is enabled whenever
the clinician removes the handpiece from the
handpiece holder.
NOTE: To remove mains power from the scaler, unplug the
power cord from the dental office power outlet.`
ON/OFF Function
The Ultrasonics is switched ON and OFF
using the treatment unit's footswitch.
Power Adjustment
Turn knob to select ultrasonic
power level for operation:
clockwise increases system
power, counter clockwise
decreases system power.
The Blue Zone™ is an extended
low-power range providing effective
subgingival debridement and greater
patient comfort during definitive therapy.
Lavage Flow Adjustment
Water flow through the handpiece is adjusted by
rotating a control element on the cable connector
(blue)
®
7.2 Steri-Mate
Handpiece
The handpiece accepts all Cavitron
Handpiece Connector
(Handpiece & Mating
Assembly are Keyed)
®
Steri-Mate
Handpiece
Insert Port
Lavage Flow Adjustment
Holding open end of handpiece upright, lavage can be increased by holding the handpiece and gently twisting the blue lavage adjustment knob
clockwise. The rate of flow through the Handpiece determines the temperature of the lavage. Low flow rates produce warmer temperatures, high
flow rates produce cooler temperatures.
ENGLISH • 10
SPS
Self Install
®
Blue Zone
®
30K™ Ultrasonic Inserts.
Cable Assembly
Date Codes (MMYY)
TYPICAL INSTALLATION
Date Codes (MMYY)
Prior to connecting, align Handpiece and Cable Assembly
electrical connections. If Cable Assembly does not seat into
Handpiece, gently rotate handpiece until contacts align.
DO NOT TWIST WHILE PUSHING
CONNECTOR INTO HANDPIECE.
BLUE Lavage Knob
loading