Cavitron
®
Ultrasonic Scaler
with Cavitron
®
Model G15
Directions For Use
Please read carefully and completely before operating unit.
Caution: United States Federal Law restricts this
device to sale by or on the order of, a licensed dental
professional.
Introduction
DENTSPLY Professional is an ISO 13485 registered
company. The DENTSPLY Cavitron
Install Scaler System is classified by Underwriters
Laboratories Inc. with respect to electric shock, fire,
and mechanical hazards in accordance with UL
60601-1 and CSA C22.2 no. 601.1.
The Cavitron SPS Self Install Ultrasonic Scaler is
equipped with a Sustained Performance System™
(SPS), which offers a constant balance between
scaling efficiency and patient comfort by maintaining
the unit's power level when the insert tip encounters
tenacious deposits allowing the clinician to effectively
scale even at a decreased/lower power setting.
The ultrasonic system consists of two parts: an insert
and the SPS
electronic system. The SPS
incorporates two closed loops. One loop provides
automatic tuning (operating frequency is adjusted
to be at resonance for each insert), the second loop
automatically controls the tip stroke over different
working conditions. The DENTSPLY Cavitron
Self Install scaler produces 30,000 microscopically
small strokes per second at the insert's working tip.
This, combined with acoustic effects of the coolant
water, produces a synergistic action that literally
"powers away" the heaviest calculus deposits while
providing exceptional operator and patient comfort.
Intended Use
This product is intended for installation into an
existing dental handpiece delivery system. The end
users are dental professionals. The devices are used
to debride light to heavy calculus deposits from the
tooth and root surfaces. Other applications specialties
include Periodontics, Orthodontics, and Endodontics.
ENGLISH • 
SPS
Self Install
Steri-Mate
Handpiece
®
®
SPS™ Self
system
®
SPS
Technical Support
For technical support and repair assistance call
DENTSPLY Professional Cavitron Care
Certified Service at 1-800-989-8826 Monday through
Friday, 8:00 A.M. to 5:00 P.M. (Eastern Time). For
areas outside the U.S., contact your local DENTSPLY
Professional representative.
Supplies & Replacement Parts
To order supplies and replacement parts in the U.S.,
contact your local DENTSPLY Professional distributor
or call 1-800-989-8826 or 717-767-8502, Monday
through Friday, 8:00 AM to 5:00 PM (Eastern Time).
For other areas, contact your local DENTSPLY
representative.
Disposal of Unit
As this is an electrical Medical Device, disposal of unit
shall be in accordance with local, federal or country
regulations.
Section 1: Indications
1.1 Ultrasonic procedures
• All general supra- and subgingival scaling
applications.
• Periodontal debridement for all types of
periodontal diseases.
• Endodontic procedures.
Section 2: Contraindications
2.1 Contraindications
Ultrasonic Systems should not be used for
restorative procedures involving condensation
of amalgam.
Section : Warnings
.1 Warnings
• Persons fitted with cardiac pacemakers,
defibrillators and other active implanted medical
devices, have been cautioned that some types
of electronic equipment might interfere with the
operation of the device. Although no instance
of interference has ever been reported to
DENTSPLY, we recommend that the handpiece
and cables be kept at least 6 to 9 inches (15 to
23 cm) away from any device and their leads
during use.
There are a variety of pacemakers and
other medically implanted devices on the
market. Clinicians should contact the device
manufacturer or the patient's physician for
specific recommendations. This unit complies
with IEC 60601 Medical Device Standards.