Section 8: Accessories
®
• Steri-Mate
Detachable Sterilizable Handpiece
• Detachable Handpiece Cable Assembly with
Swivel
®
• Cavitron
Ultrasonic Inserts
®
• Cavitron
Steri-Mate
®
• Cavitron
Steri-Mate
Kit
Section 9: Techniques for Use
9.1 Patient Positioning
• The backrest of the chair should be adjusted
for optimal access to both the upper and lower
arches. This assures patient comfort and
clinician visibility.
• Have the patient turn head to the right or left.
• Position chin up or down depending on the
quadrant and surface being treated.
• Evacuate irrigant using either a saliva ejector or
High Vacuum Evacuator (HVE).
9.2 Performing Ultrasonic Scaling Procedures
• The edges of Cavitron
intentionally rounded so there is little danger of
tissue laceration with proper ultrasonic scaling
techniques. The lips, cheek and tongue should
be retracted to prevent accidental prolonged
contact with the activated tip whenever the insert
is placed in the mouth.
• Always adjust the lavage so adequate fluid will
be available to cool the tip/tooth interface.
• In general, it is suggested that a "feather light
touch" be used both supra- and subgingivally.
The motion of the activated tip and acoustic
effects of the irrigating fluid, in most cases, is
adequate to remove even the most tenacious
calculus.
• The Blue Zone™, an extended low power range,
is marked on the dial in blue.
• Set the System's power adjustment knob to the
lowest power setting for the application and the
selected insert.
• Only adjust the system's power adjustments
knob with the insert outside the patients mouth.
9. Patient Comfort Considerations
Reasons for sensitivity
• Incorrect tip placement. Point should be directed
away from root surfaces.
• Not keeping the tip in motion on the tooth. Do
not allow the insert to remain in a static position
on any one area of the tooth. Change the insert's
path of motion.
• Applying pressure. Use extremely light grasp and
pressure, especially on exposed cementum.
• If sensitivity persists, decrease power setting
and/or move from the sensitive tooth to another
and then return.
ENGLISH • 12
®
Light Lighted Sleeve
®
Handpiece Sterile Lavage
®
Ultrasonic Inserts are
Section 10: System Care
10.1 Daily Protocol
Start-up procedures at the beginning of the day
Prior to use, inspect the Power Cord, the Scaler
housing, handpiece, and handpiece cable for damage
(nicks, cracks, or fraying). If damage is found, DO NOT
USE. Refer to the Technical Support Section for repair
service.
1. Switch ON the dental treatment unit (refer to
instruction manual of treatment unit).
2. Remove the Cavitron
assembly from handpiece holder and set
power adjustment to top of Blue Zone.
3. Connect a sterilized Steri-Mate
to the cable connector assembly. (Refer to
the Infection Control information booklet for
sterilizing instructions.)
4. Adjust the lavage control to maximum.
5. Hold the handpiece (without an insert
installed) over a sink or drain. Activate the foot
control and flush water through handpiece for
at least two minutes.
6. Place a sterilized insert into the handpiece
using a gentle push-twist motion. (Refer to the
Cavitron Ultrasonic Inserts Directions for Use
for cleaning and sterilizing instructions.)
7. Activate the ultrasonics and adjust the power
and lavage control to your preferred operating
positions.
Between patients
1. Remove any ultrasonic inserts and the
®
Steri-Mate
handpiece. Clean and sterilize
the handpiece and all inserts used during
the procedure. (Refer to the Infection
Control Information booklet and the Cavitron
Ultrasonic Inserts Directions for Use for
cleaning and sterilizing instructions.)
2. Clean and disinfect the handpiece cable
assembly by applying a water-based
disinfectant solution*, carefully following
the instructions provided by the solution
manufacturer. To clean the cable, generously
spray disinfectant solution on a clean
towel and wipe the cable and connector.
Discard used towel. To disinfect the system,
generously spray disinfectant on a clean towel
and wipe the cable and connector. Allow the
disinfectant solution to air dry.
3. Clean and disinfect all surfaces of the dental
treatment control head according to the
manufacturer's instructions.
4. Connect a sterilized Steri-Mate
onto its mating cable connector.
®
cable connector
®
handpiece
®
handpiece