Codman MedStream Manual Del Usuario página 12

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 120
Message on Control Unit
Audible Pump Alarm?
Pump Empty
Yes
Pump Battery Low
Yes
Battery Dead. Infusion
No
permanently stopped.
Explant the Pump.
Pump Clock Is Not
No
Synchronized
Pump Hardware Failure [x].
Yes
Empty the Pump Reservoir
and Contact Customer
Support.
Notes:
After you have interrogated the pump with the control unit to find out the reason for the alarm, the control unit will beep until the required action is taken.
The pump alarm will not sound if a hardware failure is detected before the preoperative internal check. The control unit will beep when the failure is detected during
pump preparation. See the MedStream Programming Guide for more information.
12
Table 3
MedStream Pump Alarms (Post-implantation)
Explanation of Condition
The volume of drug measured in the
pump reservoir is below 3 mL
Pump battery has less than 3 months
of life remaining. The pump is still
functional.
The pump battery is exhausted
The pump clock is potentially incorrect
Pump hardware failure
Required Action
Refill the pump
Schedule a pump explantation
Schedule a pump explantation; continue drug therapy via
other means of administration
If the Control Unit clock is correct, resynchronize pump
clock as prompted by the Control Unit. If Control Unit
clock is not correct, set the correct Control Unit clock,
then re-interrogate pump.
Must remove the drug from the reservoir;
call Customer Service.