EN
Steam Sterilization
Perform one of the following validated steam sterilization
cycles:
Wrapping
Method:
Dynamic Air
Cycle:
Removal
(Pre-vacuum)
Sterilization
132 °C
Temperature:
Minimum
4
Exposure
minutes
Time:
30
Minimum Dry
Time:
minutes
Immediate-Use Steam Sterilization
WARNINGS:
▪ DO NOT place equipment into a sterilization case for
immediate-use steam sterilization.
▪ After cooling to room temperature, ALWAYS use the
equipment immediately. Assurance of sterility cannot
be maintained for unwrapped equipment.
CAUTION: Stryker does not recommend immediate-use
steam sterilization for routine sterilization of medical
devices. Immediate-use steam sterilization should only
be used when individual devices require immediate
sterilization and use.
Perform one of the following validated steam sterilization
cycles:
Wrapping
Method:
Dynamic Air
Cycle:
Removal
(Pre-vacuum)
Sterilization
132 °C
Temperature:
Minimum
Exposure
4 minutes
Time:
Minimum Dry
No dry time
Time:
8
Wrapped
Gravity
134 °C
121 °C
132 °C
3
30
15
minutes
minutes
minutes
30
30
30
minutes
minutes
minutes
Unwrapped
Gravity
132 °C
10 minutes
No dry time
Ethylene Oxide (EO) Sterilization
Wrapping
Method:
Concentration:
Sterilization
Temperature:
Humidity:
Gas
Concentration:
Exposure
Time:
Minimum
Aeration Time:
Storage and Handling
WARNINGS:
▪ ALWAYS transport wrapped equipment with care to
prevent damaging the sterile barrier.
▪ ALWAYS store wrapped, processed equipment in
a controlled environment and avoid extremes in
temperature and moisture.
▪ Excessive handling of wrapped equipment will increase
the likelihood of damaging the sterile barrier and may
lead to contamination.
CAUTION: ALWAYS transport and store the equipment
within the specified environmental condition values:
Temperature
Humidity
Limitation
Limitation
NOTE: See the instructions for use supplied with the
sterilization wrap for maximum shelf-life information.
7300-001-700 Rev-B
Wrapped
100% EO
55 °C
55%
725 mg/L
80 minutes
8 hours
Atmospheric
Pressure
Limitation
www.stryker.com