EN
▪ For dead-ended lumens, once the water comes
back out of the equipment, immediately point the
distal end of the equipment down to allow the water
to drain out. Repeat this step one or two more
times until the water draining from the equipment is
clear.
▪ For all other equipment, hold the equipment on an
incline, distal end pointing down.
▪ Pay particular attention to rough surfaces, lumens,
hinges, blind holes, and joints between mating
parts.
▪ Open the door on the bottom of the handpiece
and rinse the interior of the power module
compartment.
6
▪ Actuate all moving parts to rinse hidden surfaces.
NOTE: A final rinse of the equipment using deionized or
filtered water is recommended.
6. Visually inspect the equipment for any remaining
soil or detergent solution. If soil or detergent solution
remains, repeat the cleaning procedure using fresh
detergent solution.
7. Allow the equipment to drain on absorbent wipes. If
possible, orient the equipment vertically to assist in
drainage.
8. Dry the equipment with a soft lint-free cloth, medical-
grade compressed air, or by heating the equipment in
an oven below 110 °C.
9. After manual cleaning, inspect the equipment. See the
Inspection section.
Inspection
WARNINGS:
▪ Perform recommended inspection as indicated in these
instructions.
▪ Only individuals trained and experienced in the
inspection of reusable medical devices should inspect
this equipment.
▪ DO NOT sterilize any equipment if damage is apparent
or the inspection criteria are not met.
1. After cleaning, thoroughly inspect the equipment. See
Appendix B: Inspection Criteria for inspection criteria.
2. If all of the inspection criteria are met:
▪ Charge the power modules. For charging
instructions, see the instructions for use supplied
with the battery charger.
▪ Sterilize the equipment, with the exception of the
power module. DO NOT sterilize power modules.
See the Sterilization section.
NOTE: If the equipment fails to meet the inspection
criteria, contact your Stryker sales representative or call
Stryker customer service. Outside the US, contact your
nearest Stryker subsidiary.
7300-001-700 Rev-B
www.stryker.com