1.2 EN/IEC 60601-1-2 Table 2
The Coltolux® LED unit is intended for use in the electromagnetic environment specified below . The customer or end
user of the Coltolux® LED unit should assure it is used only in such an environment .
Immunity Test
Electromagnetic
Discharge (ESD)
IEC 61000-4-2
Electrical fast
transcient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
Power Frequency
(50/60Hx) magnetic
field IEC 61000-4-8
Note UT is the a . c . mains voltage prior to the application of the test level .
Guidance and manufacturers Declaration - Electromagnetic Immunity
IEC 60601 test level
Compliance Level
+/- 8kV Contact
+/- 8kV Contact
+/- 15kV Air
+/- 15kV Air
+/- 2kV for power
+/- 2kV for power supply
supply lines
lines
+/- 1 kV for input/
+/- 1 kV for input/output
output lines
lines
+/- 1kV differential
+/- 1kV differential mode
mode (line-line)
(line-line)
+/- 2 kV common
+/- 2 kV common mode
mode (line-earth)
(line-earth)
<5% UT (>95% dip in
<5% UT (>95% dip in UT) for
UT) for 0 . 5 cycle .
0 . 5 cycle .
40% UT (60% dip in
40% UT (60% dip in UT) for
UT) for 5 cycles .
5 cycles .
70% UT (30% dip in
70% UT (30% dip in UT) for
UT) for 25 cycles .
25 cycles .
<5% UT (>95% dip in
<5% UT (>95% dip in UT) for
UT) for 5 seconds .
5 seconds .
3A/m
3A/m
Intended Electromagnetic Environ-
ment
Floors should be wood, concrete or
ceramic tile . If the floors are covered
with a synthetic material, the relative
humidity should be at least 30%
Mains power quality should be that
of a typical commercial or hospital
environment .
Mains power quality should be that
of a typical commercial or hospital
environment .
Mains power quality should be that
of a typical commercial or hospi-
tal environment . If the user of the
Coltolux® LED unit requires conti-
nued operation during power mains
interruptions, it is recommended
that the Coltolux® LED unit be powe-
red from an uninterruptable supply
with sufficient capacity to run the
unit for the maximum required time
of the interruption .
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commer-
cial or hospital environment .