VI. CLINICAL APPLICATION
Always test restorative materials with this LED
light by using the supplied test block. Failure to
do so may result in uncured material.
WARNING
Polymerization: Determine the exposure time in accordance
with the dental material manufacturer's instructions using
the test block provided . If the exposure time exceeds 20 sec,
COLTENE recommends including a cooldown time before
initiating another cycle . Failure to do so increases the risk of
tissue burns and inadvertent heating of the dental pulp .
Remove the charged handpiece from the charger base; position
the light shield; slide the single use barrier sleeve over the light;
attach the light shield (Fig. 5, Fig. 6); direct the tip of the light
lens as close to the restorative material as possible without
touching .
Activate Coltolux® LED with a momentary push of the button .
Avoid contact with un-polymerized restorative material,
glass ionomer, liquid resin or bonding agent because direct
contact will cause these materials to adhere to the light lens .
This will block the light transmission and result in improper
polymerization .
To safely terminate procedure, press and release activation
button .
VII. TROUBLESHOOTING
A.
If the light on the charger base does not illuminate:
Check to see that the power cord is securely plugged into the
charger base and the wall outlet .
Ensure the wall outlet has power by plugging a lamp into it .
B.
If the On/Off button on the handpiece does not activate the blue
LED curing light and the green light on the handpiece is not lit (and
handpiece is not in base):
The unit may be discharged and needs charging . Replace
handpiece in the charger base . If the charger base LED is orange,
handpiece is being charged . If the charger base LED is green
and the handpiece indicator light remains unlit, the unit requires
service . Contact your dealer or Coltène/ Whaledent Inc .
C.
If the On/Off button on the handpiece does not activate the curing
light and the green light above the button is illuminated, unit
requires service . Contact your dealer or Coltène/ Whaledent Inc .
D.
If your light registers a reading below 300 mW/cm
take the following steps:
Check the light lens . The light lens should be checked daily .
Replace lens with Coltène/ Whaledent Inc . Light Lens (Cat . No .
C7972) if material has cured to the lens or if it is otherwise
obstructed .
If the intensity of the light output is still under 300 mW/cm
extend the curing time according to the material's manufacturer
recommendation until the Coltolux® LED light can be serviced .
Contact your dealer or Coltène/ Whaledent Inc . for service .
E.
If activating the unit results in an error sound and the indicator is
solid orange:
The unit requires service . Contact your dealer or Coltène/
Whaledent Inc .
VIII. REPLACING THE LIGHT LENS
IMPORTANT: The Coltolux® LED Light Lens is easily scratched and
must be handled with care at all times.
WARNING
CAUTION
IX. INSPECTING THE LIGHT LENS
WARNING
X. INSPECTING THE CHARGER BASE POWER CORD
XI. CLEANING
WARNING
on a light meter,
,
2
CAUTION
Never look directly into the light emitted from
the light lens or reflected from tooth or other
surfaces. Retinal damage may occur.
Only use Coltène/ Whaledent's Light Lens (Cat.
No. C7972). Do not use other manufacturer's
light lens with this product as this will impair
performance
Do not touch the LED or reflector (Fig. 4) under
the light lens as hand oil contaminates these
components and can compromise performance.
Remove the handpiece from the charger base before
attempting to replace the light lens . Do not activate the
handpiece .
Remove the light lens by firmly pulling it from the reflector
(Fig. 4) .
Inspect the light lens before each use to insure that restorative
material has not accidentally cured to the lens . Replace lens if
material has cured to the lens or if it is otherwise obstructed .
Never look directly into the light emitted from
the light lens or reflected from tooth or other
surfaces. Retinal damage may occur.
Periodically inspect the charger base power cord for frayed areas
or broken insulation .
The below cleaning instructions are not
!
intended to replace proper infection control
procedures described in section XII of this
manual.
Do not immerse the unit in liquid, damage to the
unit will occur.
Do not attempt to clean the reflective surface
or the LED under the light lens as this will cause
damage to the unit or impair its performance.
Do not saturate the light lens with cleaning
solution or disinfectant. Do not allow solution to
remain on the light lens for an extended period of
time as this may impair performance.
Do not use petroleum-based solvents, io-
dophors, or phenolic-based products on the
handpiece or charger base. (Iodophors and
phenolics can stain the surface of the unit and
damage the plastic housing.)