ECG waveforms may have distortion of different extent. In either of the latter two modes, the system can
only show the basic ECG and the results of ST analysis may also be greatly affected. In Surgery mode,
results of ARR analysis may be somewhat affected. Therefore, it is suggested that in the environment having
relative small interference, you'd better monitor a patient in Diagnosis mode.
4 Leads of channel 2: refer to  for detailed information.
5 Waveform gain of channel 2: refer to  for detailed information.
Note
Pacemaking signal detected is marked by a "|" above the ECG waveform.

11.5 eCG Menu

11.5.1 eCG SetuP Menu
Pick the ECG hot key on the screen, and the following menu will popup.
Figure 11-6 ECG Setup Menu
ECG alarm setting
• HR ALM: pick "ON" to enable prompt message and data record during the ECG alarm; pick "OFF"
to disable the alarm function, and there will be a beside "ECG".
• ALM LEV: selectable from HIGH, MED,LOW. Level HIGH represents the most serious case.
• ALM REC: pick "ON" to enable report printing upon ECG alarm.
• ALM HI: used to set up the upper limit of ECG alarm.
• ALM LO: used to set up the lower limit of ECG alarm.
ECG alarm is activated when the heart beat exceeds set ALM HI value or falls below ALM LO value.
HR alarm range:
Patient Type
ADU
PED/NEO
Note
Please set the alarm limits according to clinical condition of individual patient. The upper limit shall not
exceed 20 beat/min higher than the patient's heart rate.
• HR FROM
ECG, SpO
, AUTO and BOTH may detect heart rate. AUTO distinguishes heart rate source according to
2
the quality of signal. By picking ECG, the monitor prompts HR and activates HR beep. By picking SpO
the monitor prompts PULSE and activates pulse beep. BOTH mode displays HR and PR simultaneously,
when this item is picked, PR parameter is displayed to the right side of SpO
in BOTH mode, HR is given the priority, i.e., if HR is available, whose sound will be sent out, but if HR is not
available, then the sound will be for PR.
• HR CHANNEL
"CH1"
to count the heart rate by CH 1 waveform
"CH2"
to count the heart rate by CH 2 waveform
"AUTO"
the monitor selects a channel automatically
MAX. ALM HI
300
350
MIN. ALM LO
15
15
. As for the sound of HR or PR
2
54
Step
1
1
,
2