Tables below describe the possible physiological alarms, technical alarms and prompt messages during ST
measurement.
Physiological alarms:
Message
Cause
ST1 TOO HIGH
ST measuring value of channel 1 is above the upper alarm limit.
STI TOO LOW
ST measuring value of channel 1 is below the lower alarm limit.
ST2 TOO HIGH
ST measuring value of channel 2 is above the upper alarm limit.
ST2 TOO LOW
ST measuring value of channel 2 is below the lower alarm limit.
Technical alarms:
Message
ST ALM LMT ERR
Prompt messages (include general alerts):
Message
Cause
ST1 EXCEED
ST measuring value of channel 1 exceeds
the measurement range.
ST2 EXCEED
ST measuring value of channel 2 exceeds
the measurement range.

11.8 Arr. Monitoring

11.8.1 Arrhythmia Analysis
The arrhythmia algorithm is used to monitor ECG of neonate and adult patient in clinical, detect the changing
of heart rate and ventricular rhythm, and also save arrhythmia events and generate alarming information.
Arrhythmia algorithm can monitor paced and non-paced patients. Qualified personnel can use arrhythmia
analysis to evaluate patient's condition (such as heart rate, PVCs frequency, rhythm and ectopic beat) and
decide the treatment. Besides detecting changing of ECG, arrhythmia algorithm can also monitor patients
and give proper alarm for arrhythmia.
• The arrhythmia monitoring is shutoff by default. You can enable it when necessary.
• This function can call up the doctor's attention to the patient's heart rate by measuring and classifying
the arrhythmia and abnormal heart beat and triggering the alarm.
• The monitor can conduct up to 13 different arrhythmia analyses.
• The monitor can store the latest 60 alarm events when taking arrhythmia analysis to a peculiar buffer.
The operator can edit these arrhythmia events through the menu below.
Pick the item ARR ANALYSIS in ECG SETUP menu to access the ARR ANALYSIS sub-menu.
ARR ANALYSIS Menu
Figure 11-11 ARR Analysis Menu
Cause
Functional safety failure
Alarm Level
Remedy
HIGH
Stop using ST alarming
function, notify biomedical
engineer or Our service staff.
Alarm Level
User-selectable
User-selectable
User-selectable
User-selectable
Alarm Level
HIGH
HIGH