14.4 Care and Cleaning

Warning
Before cleaning the monitor or the probe, make sure that the equipment is switched off and disconnected
from the power line.
Reusable TEMP Probes
1 The TEMP probe should not be heated above 100°C (212°F). It should only be subjected briefly to
temperatures between 80°C (176°F) and 100°C (212°F).
2 The probe must not be sterilized in steam.
3 Only detergents containing no alcohol can be used for disaffection.
4 The rectal probes should be used, if possible, in conjunction with a protective rubber cover.
5 To clean the probe, hold the tip with one hand and with the other hand rubbing the probe down in the
direction of the connector using a moist lint-free cloth.
Note
• Disposable TEMP probe must not be re-sterilized or reused.
• For protecting environment, the disposable TEMP probe must be recycled or disposed of properly.
Chapter 15
IBP MonItorInG(oPtIonAl)

15.1 Introduction

The Monitor measures direct blood pressure (SYS, DIA and MAP) of one selected blood vessel through two
channels, and displays two BP waveforms measures direct blood pressure (SYS, DIA and MAP).
The available pressure labels are:
Label
Definition
ART
Arterial Blood Pressure
PA
Pulmonary Arterial Pressure
CVP
Center Venous Pressure
RAP
Right Atrial Pressure
LAP
Left Atrial Pressure
ICP
Intracranium Pressure
P1-P2
Expand Pressure

15.2 Precautions during IBP Monitoring

Warning
• The operator should avoid contact with the conductive parts of the appurtenance when being connected
or applied.
• When the monitor is used with HF surgical equipment, the transducer and the cables must be avoided
conductive connection to the HF equipment to protect against burns to the patient.
• Disposable IBP transducer or domes should not be reused.
• Verify transducer cables fault detection before beginning of monitoring phase. Unplug the transducer of
the channel 1 from the socket, the screen will display the error message "IBP: SENSOR 1 OFF" and the
audible alarm is activated. The other channel is the same.
• If any kind of liquid, other than solution to be infused in pressure line or transducer, is splashed on the
equipment or its accessories, or may enter the transducer or the monitor, contact the Hospital Service
Center immediately.
Note
• Use only the pressure transducer listed in the Chapter Accessories and Ordering Information.
• The specified transducer is designed to have the special ability to protect against the electricity shock
(especially for the leak current allowed), and it is protected against the effects of a discharge of a cardiac
defibrillator. It can be used in the surgical operation. When the patient is in the defibrillation, the waveform
80