1.2 Screen display
The display of the multi-parameter monitor is a color LCD, which can display the collected patient
parameters, waveforms, alarm information as well as bed number, time and monitor status, etc.
The screen is divided into three areas (Figure 1-1): Information area 1, 4; waveform area 2; parameter area 3.
Figure 1-1 Main Display
Information Area
The Message Area is at the top and bottom part of the screen, displaying the current status of both the
monitor and the patient.
At the top (1)
Information will appear and disappear together with the reported status. According to the content, the
information is divided into:
• Battery indicator
• Filter way
• Prompt information, reporting the current status of the monitor or sensor/probe.
• flag for alarm PAUSE. Press "SILENCE" button once (less than 1 second) to mute all alarm sounds
and the flag appears at the same time. Press the button again to terminate the PAUSE status. The duration
for PAUSE status can be 1 minute, 2 minutes or 3 minutes.
• flag for alarm SILENCE. Press "SILENCE" button once (more than 1 second) to manually mute all
the sounds and this flag appears at the same time. The SILENCE status terminates when you discharge
the status or new alarm occurs.
• flag for Alarm Volume Off. It appears indicating that you have closed the alarm sound permanently.
This status terminates when you discharges the status.
Note
• If symbol appears, the system will no longer give audible alarm sound. You must be very careful
in using this function. Two ways can be used to discharge this status. One is to set the alarm volume to
an option other than OFF in the USER MAINTAIN menu. The other method is to press SILENCE button to
make the flag turn to . And then press SILENCE again and the system will restore the normal alarm
status.
• Parameter alarm information is displayed always in the upper right corner of the screen.
• When the waveforms on the screen are frozen, the FREEZE prompt will appear in the bottom part of the
screen.