6.3 recording Startup

You can start the recording in the following ways:
Continuous real-time recording
8 second real-time recording
Auto recording
Alarm recording
Frozen waveform recording
Trend graph recording
Trend table recording
Arrhythmia review recording
Alarm review recording
NIBP review recording
Titration table recording
Note
• You can press REC/STOP button on the recorder to stop the current recording process.
• Access the "RECORD" menu from the "SYSTEM SETUP" menu. Then pick the "CLEAR REC TASK" button
to stop all recording tasks, and clear all store of alarm.

6.4 recorder operations and Status Messages

Record Paper Requirement
Only standard 50 (+0/-1) mm thermosensitive record paper can be used, otherwise the recorder may not
function, the recording quality may be poor, and the thermosensitive printhead may be damaged.
Function Properly
• When the recorder is working, the record paper goes out steadily. Do not pull the paper, or the recorder
will be damaged.
• Do not operate the recorder without record paper.
Paper Out
When "RECORDER OUT OF PAPER" alarm is displayed, the recorder cannot start. Please insert record
paper properly.
Press REC/STOP to start/stop the recording.
Press REC/STOP to start recording. It will automatically stop
in 8 seconds.
Record the two waveforms selected in RECORD menu according to
the setup time interval in RECORD menu.
When alarm recording is set ON, it automatically starts when alarm
occurs.
---After accessing FREEZE menu, use knob to select two waveforms
to be output. Then press REC button in the menu to print out the
waveforms.
If two waveforms are off, the measure parameters in frozen are printed
out only.
Pick "REC" button in the "TREND GRAPH" menu when viewing
the trend graph to print out the currently displayed tr-end graph.
Pick "REC" button in the "TREND TABLE" menu when viewing
the trend table to printout the currently displayed trend table.
Access ARR RECALL window from ARR ANALYSIS of ECG SETUP
menu and Pick "WAVE" button to access "ARR WAVE RECALL" menu.
Then press "REC" button to output the Arr. Waveform and related
information currently displayed on the screen.
Access the "ALM RECALL" window from "ALARM RECALL TIME"
menu from "SYSTEM MENU" and pick "REC" button to print out
the alarm review waveform and related information currently displayed
in the "ALARM RECALL" window.
Access the "NIBP RECALL" window from "SYSTEM MENU" and pick
"REC" button to print out the NIBP information currently displayed
in the window.
Access the "DRUG CALC" menu from the "SYSTEM MENU" menu.
Pick the "TITRATION" button in the menu to access the "TITRATION"
window. Pick the "REC" button to print out the titration currently
displayed in the window.