17
3.4.1 Face Select
Select the [FACE SELECT] item in the "SYSTEM SETUP" menu to call up the following menu:
Figure 3-7 Face Select
There are five available options:
1.STANDARD SCREEN
The standard screen is the default screen. If the current screen is not the standard screen, you may enter the
standard screen by selecting STANDARD SCREEN and then selecting EXIT in FACE SELECT menu.
Figure 3-8 Standard
2.OxyCRG SCREEN
OxyCRG screen is located at the lower part of the waveform area, consisting of the HR trend, the SpO
trend, and the RR (respiration rate) trend or the compressed respiration waveform. Below the RR trend
or the compressed respiration waveform is the scale of the trend time. In addition, three labels are displayed
beneath the time scale. The labels are detailed as below.
1. Trend length
This label allows you to select the time duration of the trend graphs displayed. You can select either
1 MIN, 2 MIN or 4 MIN.
2. Compressed respiration waveform/RR trend
With this label, you can select to display the compressed respiration waveform or the RR trend beneath
the SpO
trend.
2
3. Recording
You can select the REC label to print out the the trends or the waveform displayed in the oxyCRG screen
using the recorder.
2