Figure 15-3 IBP SETUP Menu
The items to be set up in the menu include:
• ALM: Select "ON" to enable alarm prompt and data storage during IBP alarm. Select "OFF" to disable
audio alarm and prompt the symbol beside "IBP" numeric.
• ALM LEV: used to set up the alarm level. Three levels are available: HI, MED, LO.
• ALM REC: Select "ON" to enable recording during the IBP alarm or to OFF to disable the alarm recording
function.
• SWEEP: used to select the scanning speed of the IBP wave. Two selections are available: 12.5 mm/s
or 25 mm/s.
• IBP1 UNIT/IBP2 UNIT: used to select the pressure unit (mmHg / kPa / cmH2O).
• FILTER:non filter,smooth,normal.
• ALM LIMIT SET: used to access the sub-menu of IBP ALM LIMIT SET, in which the user may set up the
upper and lower alarm limit of systolic pressure, diastolic pressure and mean pressure respectively for
channel 1 and channel 2.
• SCALE ADJUST: used to access the sub-menu of IBP PRESS RULER ADJUST, in which the user may
adjust the position of the high, reference and low scales for the two waveforms displayed on the screen.
• EXPAND PRESSURE: used to access the sub-menu of IBP EXPAND PRESS SET, in which the user may
select the pressure name to be represented by P1, P2.
• DEFAULT: pick this item to access the IBP DEFAULT CONFIG dialog box, in which the user may select
whether the FACTORY DEFAULT CONFIG or the USER DEFAULT CONFIG is to be used. After selecting
any of the items and exiting the dialog box, the system will pop up the dialog box asking for the user's
confirmation.
• EXIT: used to exit the menu and return to the main screen.
Warning
• Before set the alarm limits, confirm to choose the correct label.
Figure 15-4 IBP ALM LIMIT SETUP
82