When parameter alarm occurs, two recorded waveforms can be printed out.
In order to avoid repeated printout of alarm waveforms:
• If more than two parameter alarms are switched on and triggered simultaneously, the recorder will print out
those of the highest level. If of the same alarm level, the latest alarm will be printed out.
• If an alarm occurs during the alarm of another parameter, it will be printed out after the current recording is
finished.
• If many alarms occur at the same time, some of waveforms will be stored for printout in turn.
ST Segment Alarm
The monitor records 2-channel ECG waveforms 4, 8, or 16 seconds prior to and after the alarm (totally 8, 16,
or 32 seconds) (which can be selected in the ECG SETUP menu). All parameter values during the alarm will
also be recorded.
Arrhythmia Alarm
The monitor records 2-channel ECG waveforms 4 seconds prior to and after the alarm (totally 8 seconds).
All measurement results during the alarm will also be recorded.
Freeze Waveform Recording
The monitor prints out the selected waveforms under the FREEZE mode. In this way you can snap the
abnormal waveforms on the screen and record it.
Trend Graph / Table Recording
The monitor can print out the trend graph and table in the current TREND GRAPH or TREND TABLE window.
Arrhythmia Review Recording
The monitor can print out the alarm Arrhythmia event in the current ARR RECALL window.
Alarm Review Recording
The monitor can print out the alarm events include waves and parameters in the current ALARM RECALL
window.
NIBP Review Recording
The monitor can print out all the NIBP review events in NIBP RECALL window.
Titration Table
The monitor can print out the messages in the current TITRATION window.
Notes on Recording
• Recording texts:
Real time Report
Periodic Report
Para Alarm Report: XXX (name of the alarm parameter)
Arrhythmia Report: XXX (Arrhythmia type)
Freeze Wave Report
Trend Graph
Trend Table
Para Alarm Review
NIBP Test Review
Titration Table
• Alarm parameters, alarm time and freeze time
• Patient bed number,sex, height, weight, date of birth, admission date
• Parameter name and value
• Recording time
• Waveform name
• Waveform scale (for ECG waveform)
• ECG lead, scale, filter mode, (if having ECG waveforms, it will be printed out within the first second
or when changing the lead, gain and filter mode during real-time recording.)
• Date and time
32