Gima CMS8000 Manual Del Uso página 225

Ocultar thumbs Ver también para CMS8000:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

105
"PNP"
"PNC"
"ECG LEAD OFF"
"ECG V LEAD OFF";
"ECG LL LEAD OFF";
"ECG LA LEAD OFF";
"ECG RA LEAD OFF";
"ECG C LEAD OFF";
"ECG F LEAD OFF";
"ECG L LEAD OFF";
"ECG R LEAD OFF";
SpO2 SENSOR OFF
SpO
INIT ERR
2
SpO
INIT ERR 1
2
SpO
INIT ERR 2
2
SpO
INIT ERR 3
2
SpO
INIT ERR 4
2
SpO
INIT ERR 5
2
SpO
INIT ERR 6
2
SpO
INIT ERR 7
2
SpO
INIT ERR 8
2
SpO2 COMM STOP
SpO2 COMM ERR
SpO2 ALM LMT ERR
PR ALM LMT ERR
The pacemaker is not paced.
No pacemaker signal is captured.
ECG lead is not connected correctly.
The V lead wire of ECG is not con-
nected correctly.
The LL lead wire of ECG is not con-
nected correctly.
The LA lead wire of ECG is not con-
nected correctly.
The RA lead wire of ECG is not con-
nected correctly.
The C lead wire of ECG is not con-
nected correctly.
The F lead wire of ECG is not con-
nected correctly.
The L lead wire of ECG is not con-
nected correctly.
The R lead wire of ECG is not con-
nected correctly.
SpO2 sensor may be disconnected
from the patient or the monitor.
SpO2 module failure
SpO2 module failure or communica-
tion error
SpO2 module failure or communica-
tion error
Functional safety failure
Functional safety failure
Check the connection of the pacemaker.
Check the connection of electrodes and
lead wires. Check the current situation of
the patient.
Check the connection of the pacemaker.
Check the connection of electrodes and
lead wires. Check the current situation of
the patient.
Check the connection of ECG lead wire.
Check the connection of V lead wire.
Check the connection of LL lead wire.
Check the connection of LA lead wire.
Check the connection of RA lead wire.
Check the connection of C lead wire.
Check the connection of F lead wire.
Check the connection of L lead wire.
Check the connection of R lead wire.
Make sure that the monitor and the
patient are in correct connection with the
cables.
Stop using the measuring function of
SpO2 module, notify biomedical engineer
or our service staff.
Stop using the measuring function of
SpO2 module, notify biomedical engineer
or our service staff.
Stop using the measuring function of
SpO2 module, notify biomedical engineer
or our service staff.
Stop using the measuring function of
SpO2 module, notify biomedical engineer
or our service staff.
Stop using the measuring function of
SpO2 module, notify biomedical engineer
or our service staff.
Tabla de contenido
loading

Tabla de contenido