13.5 Maintenance and Cleaning

Warning
• Do not squeeze the rubber tube on the cuff.
• Do not allow liquid to enter the connector socket at the front of the monitor.
• Do not wipe the inner part of the connector socket when cleaning the monitor.
• When the reusable cuff is not connected with the monitor, or being cleaned, always place the cover on the
rubber tube to avoid liquid permeation.
Reusable Blood Pressure Cuff
The cuff can be sterilized by means of conventional autoclaving, gas, or radiation sterilization in hot air ovens
or disinfected by immersion in decontamination solutions, but remember to remove the rubber bag if you use
this method. The cuff should not be dry-cleaned.
The cuff can also be machine-washed or hand-washed, the latter method may prolong the service life of the
cuff. Before washing, remove the latex rubber bag, and for machine-washing, close the Velcro fastening.
Allow the cuff to dry thoroughly after washing, then reinsert the rubber bag.
Figure 13-5 Replace Rubber Bag in Cuff
To replace the rubber bag in the cuff, first place the bag on top of the cuff so that the rubber tubes line up
with the large opening on the long side of the cuff. Now roll the bag lengthwise and insert it into the opening
on the long side of the cuff. Hold the tubes and the cuff and shake the complete cuff until the bag is in
position. Thread the rubber tubes from inside the cuff, and out through the small hole under the internal flap.
Disposable Blood Pressure Cuffs
Disposable cuffs are intended for one-patient use only. Do not use the same cuff on any other patient.
Do not sterilize or use autoclave on disposable cuffs. Disposable cuffs can be cleaned using soap solution
to prevent infection.
Note
For protecting environment, the disposable blood pressure cuffs must be recycled or disposed of properly.
Cause
During manual measuring mode.
During continuous measuring mode.
During automatic measuring mode.
After selecting interval between measurements in MENU
Press START/STOP key during measuring to stop measurement
During calibrating
Calibration over
During pneumatic test
pneumatic test over
NIBP module in resetting
NIBP module reset failed
Alarm Level
No alarm